21 CFR 1020.40(c)(6)(iii): Cabinet x-ray systems

PART 1020-PERFORMANCE STANDARDS FOR IONIZING RADIATION EMITTING PRODUCTS

Key facts

Citation 21 CFR 1020.40(c)(6)(iii)
Section Cabinet x-ray systems
Part PART 1020-PERFORMANCE STANDARDS FOR IONIZING RADIATION EMITTING PRODUCTS
Subchapter 21 CFR J Radiological
Product types cited Devices
Regulation text (excerpt) (iii) Two independent means which indicate when and only when x-rays are being generated, unless the x-ray generation period is less than one-half second, in which case the indicators shall be activated for one-half second, and which are discernible from any point at which initiation of x-ray generation is possible. Failure of a single component of the cabinet x-ray system shall not cause failure of both indicators to perform their intended function. One, but not both, of the indicators required…