21 CFR 1020.30(k): Diagnostic x-ray systems and their major components

PART 1020-PERFORMANCE STANDARDS FOR IONIZING RADIATION EMITTING PRODUCTS

Key facts

Citation 21 CFR 1020.30(k)
Section Diagnostic x-ray systems and their major components
Part PART 1020-PERFORMANCE STANDARDS FOR IONIZING RADIATION EMITTING PRODUCTS
Subchapter 21 CFR J Radiological
Product types cited Devices
Regulation text (excerpt) (k) Leakage radiation from the diagnostic source assembly. The leakage radiation from the diagnostic source assembly measured at a distance of 1 meter in any direction from the source shall not exceed 0.88 milligray (mGy) air kerma (vice 100 milliroentgen (mR) exposure) in 1 hour when the x-ray tube is operated at the leakage technique factors. If the maximum rated peak tube potential of the tube housing assembly is greater than the maximum rated peak tube potential for the diagnostic source…