21 CFR 1020.30(k): Diagnostic x-ray systems and their major components
PART 1020-PERFORMANCE STANDARDS FOR IONIZING RADIATION EMITTING PRODUCTS
Key facts
| Citation | 21 CFR 1020.30(k) |
|---|---|
| Section | Diagnostic x-ray systems and their major components |
| Part | PART 1020-PERFORMANCE STANDARDS FOR IONIZING RADIATION EMITTING PRODUCTS |
| Subchapter | 21 CFR J Radiological |
| Product types cited | Devices |
| Regulation text (excerpt) | (k) Leakage radiation from the diagnostic source assembly. The leakage radiation from the diagnostic source assembly measured at a distance of 1 meter in any direction from the source shall not exceed 0.88 milligray (mGy) air kerma (vice 100 milliroentgen (mR) exposure) in 1 hour when the x-ray tube is operated at the leakage technique factors. If the maximum rated peak tube potential of the tube housing assembly is greater than the maximum rated peak tube potential for the diagnostic source… |