21 CFR 1020.40(c)(4)(i): Cabinet x-ray systems

PART 1020-PERFORMANCE STANDARDS FOR IONIZING RADIATION EMITTING PRODUCTS

Key facts

Citation 21 CFR 1020.40(c)(4)(i)
Section Cabinet x-ray systems
Part PART 1020-PERFORMANCE STANDARDS FOR IONIZING RADIATION EMITTING PRODUCTS
Subchapter 21 CFR J Radiological
Product types cited Devices
Regulation text (excerpt) (i) Each door of a cabinet x-ray system shall have a minimum of two safety interlocks. One, but not both of the required interlocks shall be such that door opening results in physical disconnection of the energy supply circuit to the high-voltage generator, and such disconnection shall not be dependent upon any moving part other than the door.

At a glance

  • 1–9 Inspections citing this section
  • None Warning letters citing this section

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About CFR sections cited by the FDA

Title 21 of the Code of Federal Regulations contains the FDA's regulations, including current good manufacturing practice and quality system requirements that inspections cite.

Source: U.S. Code of Federal Regulations (eCFR); citation statistics from FDA inspection and warning-letter data