International safety notice

The lithium-ion battery for the autopulse artificial resuscitation system has been fully charged overseas. Despite this, there have been reports of the device not starting up. As a result of the investigation, it was determined that there is a possibility that this event may occur due to the semiconductor element used on the printed circuit board of the product with a specific serial number. As we have discovered that there is a problem, we are implementing a voluntary recall of the affected products.

Japan / PMDA Recalls · JPN · Sep 26, 2022

Key facts

Title The lithium-ion battery for the autopulse artificial resuscitation system has been fully charged overseas. Despite this, there have been reports of the device not starting up. As a result of the investigation, it was determined that there is a possibility that this event may occur due to the semiconductor element used on the printed circuit board of the product with a specific serial number. As we have discovered that there is a problem, we are implementing a voluntary recall of the affected products.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: Electric cardiopulmonary resuscitator Sales name: Autopulse artificial resuscitation system Model 100 V1.5
Manufacturer Asahi Kasei Sol Medical Co., Ltd.
Models 08647, 8700-0752-03, 08606, 08640
Date Sep 26, 2022
Regulator page info.pmda.go.jp

At a glance

  • No Recall
  • No Field safety notice

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About International device safety notices

Field safety notices, recalls and safety alerts published by about 30 national medical device regulators outside the United States, matched by Orca1 to FDA device data.

Source: National medical device regulators (as listed), compiled by Orca1

Browse all International device safety notices