International safety notice

When a medical institution conducted a "battery inspection" on this product, it was found that the "battery has deteriorated. We have received reports that the message "Please replace" is displayed. As a result of the investigation, it was found that changing the battery may cause the capacity to be incorrectly recognized under certain usage conditions. Therefore, we are conducting a voluntary recall to replace the batteries with ones that do not cause misidentification of capacity.

Japan / PMDA Recalls · JPN · Apr 9, 2021

Key facts

Title When a medical institution conducted a "battery inspection" on this product, it was found that the "battery has deteriorated. We have received reports that the message "Please replace" is displayed. As a result of the investigation, it was found that changing the battery may cause the capacity to be incorrectly recognized under certain usage conditions. Therefore, we are conducting a voluntary recall to replace the batteries with ones that do not cause misidentification of capacity.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: Syringe infusion pump Brand name: (1) JMS syringe pump SP-520 (2) JMS syringe pump SP-120
Manufacturer JMS Co., Ltd.
Models 20041632, 19A3811B, 19E4061B, 18G2959B, 17K0496P, 20051757, 18E2776B, 18A2534B, 15120334, 18L3381B, 16J0338P, 17K0508P and 2782 more
Date Apr 9, 2021
Regulator page info.pmda.go.jp

At a glance

  • No Recall
  • No Field safety notice
  • 1–9 Matching device models in FDA data
  • 1–9 Matching catalog numbers in FDA data

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About International device safety notices

Field safety notices, recalls and safety alerts published by about 30 national medical device regulators outside the United States, matched by Orca1 to FDA device data.

Source: National medical device regulators (as listed), compiled by Orca1

Browse all International device safety notices