International safety notice

This product is used by connecting to a centrifugal pump drive device for extracorporeal circulation equipment (hereinafter referred to as the "drive device"). We received a report from a medical institution that the product was not recognized when connected to the drive device. As a result of the investigation, it was determined that the cause was that the conductive tape inside the pressure gauge had peeled off and came into contact with the board, so we conducted a voluntary recall. vinegar.

Japan / PMDA Recalls · JPN · Apr 2, 2024

Key facts

Title This product is used by connecting to a centrifugal pump drive device for extracorporeal circulation equipment (hereinafter referred to as the "drive device"). We received a report from a medical institution that the product was not recognized when connected to the drive device. As a result of the investigation, it was determined that the cause was that the conductive tape inside the pressure gauge had peeled off and came into contact with the board, so we conducted a voluntary recall. vinegar.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: Heart-lung machine pressure gauge Brand name: Artificial heart-lung pressure gauge PS-1
Manufacturer JMS Co., Ltd.
Models 24010760TN, JM-PS1-6P-TN, 23110725TN
Date Apr 2, 2024
Regulator page info.pmda.go.jp

At a glance

  • No Recall
  • No Field safety notice

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About International device safety notices

Field safety notices, recalls and safety alerts published by about 30 national medical device regulators outside the United States, matched by Orca1 to FDA device data.

Source: National medical device regulators (as listed), compiled by Orca1