International safety notice

As a result of product analysis by the overseas manufacturer, the characteristics of defibrillation pad pack HDF-PD-3150 (hereinafter referred to as pad pack) In certain lots, battery cells that were in contact with the welded part due to defective spot welding of the installed battery. Because it was discovered that short circuits could occur in batteries, leakage current could prematurely drain the battery over time. We have decided to carry out a voluntary recall to replace the pack.

Japan / PMDA Recalls · JPN · Jun 2, 2023

Key facts

Title As a result of product analysis by the overseas manufacturer, the characteristics of defibrillation pad pack HDF-PD-3150 (hereinafter referred to as pad pack) In certain lots, battery cells that were in contact with the welded part due to defective spot welding of the installed battery. Because it was discovered that short circuits could occur in batteries, leakage current could prematurely drain the battery over time. We have decided to carry out a voluntary recall to replace the pack.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: Automatic defibrillator for non-medical personnel Brand name: Components of "Automated External Defibrillator Rescue Heart HDF-3500" Defibrillation Pad Pack HDF-PD-3150
Manufacturer OMRON Healthcare Co., Ltd.
Models C0331, HDF-PD-3150, C0329, C0245, C0283, C0335, C0333, C0250, C0247, C0286, C0330, C0248 and 13 more
Date Jun 2, 2023
Regulator page info.pmda.go.jp

At a glance

  • No Recall
  • No Field safety notice
  • 1–9 Matching catalog numbers in FDA data

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About International device safety notices

Field safety notices, recalls and safety alerts published by about 30 national medical device regulators outside the United States, matched by Orca1 to FDA device data.

Source: National medical device regulators (as listed), compiled by Orca1

Browse all International device safety notices