International safety notice
Due to an inappropriate inspection method for the inspection (polarization test) conducted during the manufacturing process of the product (XFine lead), regulations were not met. There is a possibility that leads exhibiting polarization voltage exceeding the standard (highly polarized leads) may be shipped without being properly detected and eliminated. has been confirmed. The highly polarized lead is sold under the name “Alizea” (common name: implantable cardiac pacemaker, approval number: 30400BZX00001000) When connected to the MV (Minute Ventilation) sensor function, a highly polarized state at the lead tip can be detected by the pacemaker. Depending on settings such as sensitivity, oversensing may occur. Improvements have already been made to address the inappropriateness of the testing method, and no products with problems due to the current testing method have been shipped. However, since the contamination of highly polarized leads cannot be ruled out in the product before improvement, we Voluntary recall of products and regular follow-up with medical institutions regarding patients implanted with the affected products We have decided to implement this. As mentioned above, improvements to the testing method have already been made, so it is unlikely that this phenomenon will occur in future products that will be newly implanted. There is no possibility of this happening. In addition, there is no possibility that this event will occur except with specific settings for specific products.
Japan / PMDA Recalls · JPN · Aug 31, 2022
Key facts
| Title | Due to an inappropriate inspection method for the inspection (polarization test) conducted during the manufacturing process of the product (XFine lead), regulations were not met. There is a possibility that leads exhibiting polarization voltage exceeding the standard (highly polarized leads) may be shipped without being properly detected and eliminated. has been confirmed. The highly polarized lead is sold under the name “Alizea” (common name: implantable cardiac pacemaker, approval number: 30400BZX00001000) When connected to the MV (Minute Ventilation) sensor function, a highly polarized state at the lead tip can be detected by the pacemaker. Depending on settings such as sensitivity, oversensing may occur. Improvements have already been made to address the inappropriateness of the testing method, and no products with problems due to the current testing method have been shipped. However, since the contamination of highly polarized leads cannot be ruled out in the product before improvement, we Voluntary recall of products and regular follow-up with medical institutions regarding patients implanted with the affected products We have decided to implement this. As mentioned above, improvements to the testing method have already been made, so it is unlikely that this phenomenon will occur in future products that will be newly implanted. There is no possibility of this happening. In addition, there is no possibility that this event will occur except with specific settings for specific products. |
|---|---|
| Regulator | Japan / PMDA Recalls |
| Country | JPN |
| Brand | Common name: Endocardial implantable pacemaker lead Sales name: XFine Lead |
| Manufacturer | Japan Microport CRM Co., Ltd. |
| Models | TTX502005, TJX405520, TJX405458, TTX502192, TTX641555, TJX510052, TTX501793, TJX405263, TJX405226, TTX641585, TJX405097, TTX654064 and 371 more |
| Date | Aug 31, 2022 |
| Regulator page | info.pmda.go.jp |
At a glance
- No Recall
- No Field safety notice
Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.
About International device safety notices
Field safety notices, recalls and safety alerts published by about 30 national medical device regulators outside the United States, matched by Orca1 to FDA device data.
Source: National medical device regulators (as listed), compiled by Orca1