International safety notice

Due to an inappropriate inspection method for the inspection (polarization test) conducted during the manufacturing process of the product (XFine lead), regulations were not met. There is a possibility that leads exhibiting polarization voltage exceeding the standard (highly polarized leads) may be shipped without being properly detected and eliminated. has been confirmed. The highly polarized lead is sold under the name “Alizea” (common name: implantable cardiac pacemaker, approval number: 30400BZX00001000) When connected to the MV (Minute Ventilation) sensor function, a highly polarized state at the lead tip can be detected by the pacemaker. Depending on settings such as sensitivity, oversensing may occur. Improvements have already been made to address the inappropriateness of the testing method, and no products with problems due to the current testing method have been shipped. However, since the contamination of highly polarized leads cannot be ruled out in the product before improvement, we Voluntary recall of products and regular follow-up with medical institutions regarding patients implanted with the affected products We have decided to implement this. As mentioned above, improvements to the testing method have already been made, so it is unlikely that this phenomenon will occur in future products that will be newly implanted. There is no possibility of this happening. In addition, there is no possibility that this event will occur except with specific settings for specific products.

Japan / PMDA Recalls · JPN · Aug 31, 2022

Key facts

Title Due to an inappropriate inspection method for the inspection (polarization test) conducted during the manufacturing process of the product (XFine lead), regulations were not met. There is a possibility that leads exhibiting polarization voltage exceeding the standard (highly polarized leads) may be shipped without being properly detected and eliminated. has been confirmed. The highly polarized lead is sold under the name “Alizea” (common name: implantable cardiac pacemaker, approval number: 30400BZX00001000) When connected to the MV (Minute Ventilation) sensor function, a highly polarized state at the lead tip can be detected by the pacemaker. Depending on settings such as sensitivity, oversensing may occur. Improvements have already been made to address the inappropriateness of the testing method, and no products with problems due to the current testing method have been shipped. However, since the contamination of highly polarized leads cannot be ruled out in the product before improvement, we Voluntary recall of products and regular follow-up with medical institutions regarding patients implanted with the affected products We have decided to implement this. As mentioned above, improvements to the testing method have already been made, so it is unlikely that this phenomenon will occur in future products that will be newly implanted. There is no possibility of this happening. In addition, there is no possibility that this event will occur except with specific settings for specific products.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: Endocardial implantable pacemaker lead Sales name: XFine Lead
Manufacturer Japan Microport CRM Co., Ltd.
Models TTX502005, TJX405520, TJX405458, TTX502192, TTX641555, TJX510052, TTX501793, TJX405263, TJX405226, TTX641585, TJX405097, TTX654064 and 371 more
Date Aug 31, 2022
Regulator page info.pmda.go.jp

At a glance

  • No Recall
  • No Field safety notice

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About International device safety notices

Field safety notices, recalls and safety alerts published by about 30 national medical device regulators outside the United States, matched by Orca1 to FDA device data.

Source: National medical device regulators (as listed), compiled by Orca1

Browse all International device safety notices