International safety notice

The noise suppression capacitor used in the control parts was damaged. Unusual behavior was observed. Normally, the movement is proportional to the shaking of the user's body, It was found that the damage caused the device to suddenly move rapidly regardless of the shaking of the body. The load on a capacitor for noise suppression varies depending on the installation environment of the equipment. Although this damage was judged to be a rare case, Due to the risk of the user falling due to malfunction, We have decided to recall all products.

Japan / PMDA Recalls · JPN · Feb 22, 2024

Key facts

Title The noise suppression capacitor used in the control parts was damaged. Unusual behavior was observed. Normally, the movement is proportional to the shaking of the user's body, It was found that the damage caused the device to suddenly move rapidly regardless of the shaking of the body. The load on a capacitor for noise suppression varies depending on the installation environment of the equipment. Although this damage was judged to be a rare case, Due to the risk of the user falling due to malfunction, We have decided to recall all products.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: Balance function meter Sales name: BASYS
Manufacturer Tech Gihan Co., Ltd.
Models SMC10021Y056, SMB100184034, SMB100204047, 87037, SMB100172019, SMB10018102, SMB100181029, SMC100221057, SMB100196041, SMB100218054, SMB10005, SMB100178026 and 41 more
Date Feb 22, 2024
Regulator page info.pmda.go.jp

At a glance

  • No Recall
  • No Field safety notice

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About International device safety notices

Field safety notices, recalls and safety alerts published by about 30 national medical device regulators outside the United States, matched by Orca1 to FDA device data.

Source: National medical device regulators (as listed), compiled by Orca1

Browse all International device safety notices