International safety notice

We received a report of a problem that the main battery of this device was emitting smoke. As a result of the investigation, there were no similar problems occurring in other devices of the same model, and we believe that there is a high possibility that this was an individual (single) event. However, the main battery was determined to have a "defective labeling" during the manufacturer's acceptance inspection, and after being returned to the battery manufacturer, it was repaired. It has been discovered that there is a history of At this time, it has not been determined that the repair was the cause of the smoke, but we have confirmed that the main battery has a history of repair. We have decided to recall the product because we cannot deny the possibility that a similar problem may occur.

Japan / PMDA Recalls · JPN · Jun 24, 2021

Key facts

Title We received a report of a problem that the main battery of this device was emitting smoke. As a result of the investigation, there were no similar problems occurring in other devices of the same model, and we believe that there is a high possibility that this was an individual (single) event. However, the main battery was determined to have a "defective labeling" during the manufacturer's acceptance inspection, and after being returned to the battery manufacturer, it was repaired. It has been discovered that there is a history of At this time, it has not been determined that the repair was the cause of the smoke, but we have confirmed that the main battery has a history of repair. We have decided to recall the product because we cannot deny the possibility that a similar problem may occur.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: Oxygen concentrator Sales name: High Sanso Portable α
Manufacturer Teijin Pharma Co., Ltd.
Models HW208C0112, HW132X0063, HW182C0055, HW174C0010, HW208C0129, HW153C0027, HW186C0090, HW208C0144, HW182C0002, HW14BC0023, HW154C0010, HW208C0071 and 346 more
Date Jun 24, 2021
Regulator page info.pmda.go.jp

At a glance

  • No Recall
  • No Field safety notice

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About International device safety notices

Field safety notices, recalls and safety alerts published by about 30 national medical device regulators outside the United States, matched by Orca1 to FDA device data.

Source: National medical device regulators (as listed), compiled by Orca1