According to the manufacturer, when measuring creatinine with the medical device, in rare cases, the patient's measurement result will be lower than the expected value. I was informed that there was a possibility. As a result of an investigation by an overseas manufacturer, it was determined that this phenomenon was caused by the creatinine membrane, and we have now confirmed that the cause of this phenomenon has been confirmed. We have received notification that the improvement of the creatinine membrane has been completed, so we will review the creatinine membrane before the improvement. We will carry out the collection. Regarding this event, we have already provided information to the relevant facilities and thoroughly informed them about preventive measures against this event. Masu.

Japan / PMDA Recalls · JPN · May 20, 2021

Key facts

Title According to the manufacturer, when measuring creatinine with the medical device, in rare cases, the patient's measurement result will be lower than the expected value. I was informed that there was a possibility. As a result of an investigation by an overseas manufacturer, it was determined that this phenomenon was caused by the creatinine membrane, and we have now confirmed that the cause of this phenomenon has been confirmed. We have received notification that the improvement of the creatinine membrane has been completed, so we will review the creatinine membrane before the improvement. We will carry out the collection. Regarding this event, we have already provided information to the relevant facilities and thoroughly informed them about preventive measures against this event. Masu.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: General purpose blood gas analyzer Sales name: ABL800 FLEX system
Manufacturer Radiometer Co., Ltd.
Models R0994, R0987, R0996, R0984, R1020, R0997, R1018, R1001, R1022, R0988, R0998, R1008 and 22 more
Date May 20, 2021
Regulator page info.pmda.go.jp

At a glance

  • No Recall
  • No Field safety notice
  • 1–9 Matching device models in FDA data

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About International device safety notices

Field safety notices, recalls and safety alerts published by about 30 national medical device regulators outside the United States, matched by Orca1 to FDA device data.

Source: National medical device regulators (as listed), compiled by Orca1

Browse all International device safety notices