International safety notice
During internal confirmation testing, due to a problem with the built-in software, air leakage occurred during operation at a set speed of 1 to 2mL/h and during KVO operation. It turned out that the foam alarm did not go off. We will carry out a voluntary recall of the product.
Japan / PMDA Recalls · JPN · May 21, 2024
Key facts
| Title | During internal confirmation testing, due to a problem with the built-in software, air leakage occurred during operation at a set speed of 1 to 2mL/h and during KVO operation. It turned out that the foam alarm did not go off. We will carry out a voluntary recall of the product. |
|---|---|
| Regulator | Japan / PMDA Recalls |
| Country | JPN |
| Brand | Common name: General purpose infusion pump Brand name: Top infusion pump TOP-i200 |
| Manufacturer | Top Co., Ltd. |
| Models | VT00539, VT00197, VT00558, VT00200, VT00703, VT00103, VT00286, VT00559, VT00545, VT00756, VT00527, VU01089 and 38 more |
| Date | May 21, 2024 |
| Regulator page | info.pmda.go.jp |