International safety notice

During internal confirmation testing, due to a problem with the built-in software, air leakage occurred during operation at a set speed of 1 to 2mL/h and during KVO operation. It turned out that the foam alarm did not go off. We will carry out a voluntary recall of the product.

Japan / PMDA Recalls · JPN · May 21, 2024

Key facts

Title During internal confirmation testing, due to a problem with the built-in software, air leakage occurred during operation at a set speed of 1 to 2mL/h and during KVO operation. It turned out that the foam alarm did not go off. We will carry out a voluntary recall of the product.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: General purpose infusion pump Brand name: Top infusion pump TOP-i200
Manufacturer Top Co., Ltd.
Models VT00539, VT00197, VT00558, VT00200, VT00703, VT00103, VT00286, VT00559, VT00545, VT00756, VT00527, VU01089 and 38 more
Date May 21, 2024
Regulator page info.pmda.go.jp