When importing measurement data from an ultrasound diagnostic device with the program medical device, if conditions overlap, data from another patient may be It has been found that there is a possibility that the data may be mixed with the measured data of. For this reason, we will replace the entire medical device program that corrected the defect and reinstall it as a recall. I will do it.

Japan / PMDA Recalls · JPN · Jun 8, 2023

Key facts

Title When importing measurement data from an ultrasound diagnostic device with the program medical device, if conditions overlap, data from another patient may be It has been found that there is a possibility that the data may be mixed with the measured data of . For this reason, we will replace the entire medical device program that corrected the defect and reinstall it as a recall. I will do it.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: Program for ultrasound machine workstations Sales name: Ultrasound image analysis program UltraExtend NX CUW-U001S
Manufacturer Canon Medical Systems Inc.
Models 99A2292006, CUW-U001S, 99A22Z2010, 99A2222001, 99A2222005, 99A2292007
Date Jun 8, 2023
Regulator page info.pmda.go.jp