International safety notice

We have received information from a customer that the sterile packaging of the product in question was found to be unsealed. When we investigated the actual product, we found that an error had occurred in the sterilization packaging sealing process and was overlooked during the inspection. It turned out that it had been leaked. We have decided to carry out a voluntary recall for Lot.

Japan / PMDA Recalls · JPN · Feb 24, 2022

Key facts

Title We have received information from a customer that the sterile packaging of the product in question was found to be unsealed. When we investigated the actual product, we found that an error had occurred in the sterilization packaging sealing process and was overlooked during the inspection. It turned out that it had been leaked. We have decided to carry out a voluntary recall for Lot.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: Infusion accessory set Sales name: Closed C
Manufacturer Fortegrow Medical Co., Ltd.
Models 1880-FCC, 211102
Date Feb 24, 2022
Regulator page info.pmda.go.jp

At a glance

  • No Recall
  • No Field safety notice

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About International device safety notices

Field safety notices, recalls and safety alerts published by about 30 national medical device regulators outside the United States, matched by Orca1 to FDA device data.

Source: National medical device regulators (as listed), compiled by Orca1