International safety notice

The concentration of the standard gas used for calibration was outside the standard value range. Measurement results that exceed the error range of nitric oxide (NO) concentration measurement accuracy set in the approval document are displayed. An internal investigation revealed this possibility. Therefore, we calibrated using a standard gas that was outside the standard value range. We will collect the shipped product with the relevant serial number.

Japan / PMDA Recalls · JPN · Jun 17, 2024

Key facts

Title The concentration of the standard gas used for calibration was outside the standard value range. Measurement results that exceed the error range of nitric oxide (NO) concentration measurement accuracy set in the approval document are displayed. An internal investigation revealed this possibility. Therefore, we calibrated using a standard gas that was outside the standard value range. We will collect the shipped product with the relevant serial number.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: Nitric oxide gas analyzer Brand name: (1) Breath NO monitor NObreath (2) Breath NO monitor NObreath V2
Manufacturer Harada Sangyo Co., Ltd.
Models NN006059, NN006056, NN006052, NBR025768, NN006054, NBR025069, NBR024405, NBR024159, NN006060, NN006062, NN006057, NBR025601 and 2 more
Date Jun 17, 2024
Regulator page info.pmda.go.jp

At a glance

  • No Recall
  • No Field safety notice

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About International device safety notices

Field safety notices, recalls and safety alerts published by about 30 national medical device regulators outside the United States, matched by Orca1 to FDA device data.

Source: National medical device regulators (as listed), compiled by Orca1

Browse all International device safety notices