International safety notice

Software version 1.1.0. 1: The measurement tool has a maximum error of 14%. 2: If there is a partial space in the puncture window of the compression plate, problems may occur in image processing. These issues do not affect the safety or effectiveness of the system, but the bugs have been fixed in Software 1.1.3. We will carry out a voluntary renovation to introduce 0.

Japan / PMDA Recalls · JPN · Nov 14, 2022

Key facts

Title Software version 1.1.0. 1: The measurement tool has a maximum error of 14%. 2: If there is a partial space in the puncture window of the compression plate, problems may occur in image processing. These issues do not affect the safety or effectiveness of the system, but the bugs have been fixed in Software 1.1.3. We will carry out a voluntary renovation to introduce 0.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: Stationary digital breast X-ray diagnostic device Brand name: Affirm Prone Biopsy System
Manufacturer Hologic Japan Co., Ltd.
Models APB1100982, APB1100894, APB1100829, APB1100880, APB1101002, APB1100914, APB1100946, APB1100942, 100956, APB1100907, APB1101003, APB1101006 and 2 more
Date Nov 14, 2022
Regulator page info.pmda.go.jp

At a glance

  • No Recall
  • No Field safety notice

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About International device safety notices

Field safety notices, recalls and safety alerts published by about 30 national medical device regulators outside the United States, matched by Orca1 to FDA device data.

Source: National medical device regulators (as listed), compiled by Orca1

Browse all International device safety notices