International safety notice

When setting the flow rate including the first decimal point, the value after the decimal point is not reflected, resulting in underdosing. It was discovered that there was a firmware defect in the control CPU. Therefore, the firmware of the devices subject to recall will be updated to firmware that has been addressed. We will carry out a voluntary recall. Please note that this issue only applies to devices equipped with a dedicated mode.

Japan / PMDA Recalls · JPN · Nov 10, 2023

Key facts

Title When setting the flow rate including the first decimal point, the value after the decimal point is not reflected, resulting in underdosing. It was discovered that there was a firmware defect in the control CPU. Therefore, the firmware of the devices subject to recall will be updated to firmware that has been addressed. We will carry out a voluntary recall. Please note that this issue only applies to devices equipped with a dedicated mode.
Regulator Japan / PMDA Recalls
Country JPN
Brand Common name: General purpose infusion pump Brand name: CureSense infusion pump IP-100
Manufacturer Metek Co., Ltd.
Models 21I0596D1, 21G0479D1, 20J0069D1, 20J0067D1, 21B0185D1, 21J0646D1, 21E0385D1, 20J0029D1, 22I0115S1, 21F0445D1, 21G0450D1, 22I0086S1 and 167 more
Date Nov 10, 2023
Regulator page info.pmda.go.jp

At a glance

  • No Recall
  • No Field safety notice

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About International device safety notices

Field safety notices, recalls and safety alerts published by about 30 national medical device regulators outside the United States, matched by Orca1 to FDA device data.

Source: National medical device regulators (as listed), compiled by Orca1

Browse all International device safety notices