Drug recall

VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006

FDA recall event 99704 · Class II · Steris Corporation

Key facts

Event ID 99704
Recall numbers Z-0070-2027, Z-0071-2027, Z-0072-2027, Z-0073-2027, Z-0074-2027, Z-0075-2027
Recalling firm Steris Corporation
Product type Devices
Products VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006, VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007, VERIFY Biological Steam Test Pack Model/Catalog Number: S3065, VERIFY Biological Steam Test Pack Model/Catalog Number: S3069, VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060 and 1 more
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the…
Initiation date Aug 31, 2026
Report date Oct 14, 2026
City Mentor
State OH
Country United States
FEI number 1527821