Drug recall

Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

FDA recall event 99331 · Class II · Daavlin Distributing Company

Key facts

Event ID 99331
Recall numbers Z-2742-2026, Z-2743-2026, Z-2744-2026
Recalling firm Daavlin Distributing Company
Product type Devices
Products Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders, Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders, Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Initiation date Mar 5, 2026
Report date Jul 22, 2026
City Bryan
State OH
Country United States
FEI number 1526255