Drug recall

Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator

FDA recall event 99075 · Class II · Accuray Incorporated

Key facts

Event ID 99075
Recall numbers Z-2640-2026
Recalling firm Accuray Incorporated
Product type Devices
Products Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using…
Initiation date Jun 16, 2026
Report date Jul 8, 2026
City Madison
State WI
Country United States
FEI number 3003873069

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls