Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product…

FDA recall event 99036 · Class I · Datex-Ohmeda Inc

Key facts

Event ID 99036
Recall numbers Z-2357-2026, Z-2358-2026, Z-2359-2026
Recalling firm Datex-Ohmeda Inc
Product type Devices
Products Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 2082663-001, 2082663-002, M1241424 Product impacted only if service blender M1091607-R…, Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R…, Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.
Classification Class I
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user…
Initiation date May 8, 2026
Report date Jun 24, 2026
City Waukesha
State WI
Country United States
FEI number 3036470679

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls