Drug recall

RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U…

FDA recall event 98729 · Class I · Bolton Medical Inc.

Key facts

Event ID 98729
Recall numbers Z-2160-2026
Recalling firm Bolton Medical Inc.
Product type Devices
Products RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-34-109-34U…
Classification Class I
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.
Initiation date Apr 22, 2026
Report date Jun 3, 2026
City Sunrise
State FL
Country United States
FEI number 1000135961

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls