Philips Tempus Pro Patient Monitor, REF: 00-1004, 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R. USB…
FDA recall event 97889 · Class II · Remote Diagnostic Technologies Ltd.
Key facts
| Event ID | 97889 |
|---|---|
| Recall numbers | Z-1027-2026 |
| Recalling firm | Remote Diagnostic Technologies Ltd. |
| Product type | Devices |
| Products | Philips Tempus Pro Patient Monitor, REF: 00-1004, 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R. USB 2-Channel IBP Module, REF: 01-2017. Used with User/Operator Manual |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | Patient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for accuracy. If healthcare providers base clinical… |
| Initiation date | Nov 26, 2025 |
| Report date | Jan 14, 2026 |
| City | Farnborough |
| Country | United Kingdom |
| FEI number | 3003832357 |