Drug recall

Anesthesia circuit kit; CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230

FDA recall event 97882 · Class I · Medline Industries, LP

Key facts

Event ID 97882
Recall numbers Z-0442-2026, Z-0443-2026, Z-0444-2026, Z-0445-2026, Z-0446-2026, Z-0447-2026, Z-0448-2026, Z-0449-2026, Z-0450-2026, Z-0451-2026
Recalling firm Medline Industries, LP
Product type Devices
Products Anesthesia circuit kit; CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230, Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550, Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D, Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909116D; VASCULAR DYNJ909124F, Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES… and 5 more
Classification Class I
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of…
Initiation date Oct 10, 2025
Report date Nov 19, 2025
City Northfield
State IL
Country United States
FEI number 1417592

At a glance

  • 10–99 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls