Drug recall
IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra
FDA recall event 97750 · Class I · IMPRIS Imaging Inc
Key facts
| Event ID | 97750 |
|---|---|
| Recall numbers | Z-0161-2026, Z-0162-2026, Z-0163-2026, Z-0164-2026 |
| Recalling firm | IMPRIS Imaging Inc |
| Product type | Devices |
| Products | IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra, IMRIS Operating Suite, IMRIS iMRI 3T V, containing Siemens 3 Tesla MRI Magnet, vIDA, IMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida, IMRIS Operating Suite, Neuro III-SV iMRI System, iMRX System, containing Siemens 3 Tesla MRI Magnet, Verio |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist… |
| Initiation date | Sep 30, 2025 |
| Report date | Nov 5, 2025 |
| City | Chaska |
| State | MN |
| Country | United States |
| FEI number | 3010326005 |