Drug recall

Biograph mMR. Model Number: 10433372.

FDA recall event 97530 · Class I · Siemens Medical Solutions USA, Inc

Key facts

Event ID 97530
Recall numbers Z-2654-2025, Z-2655-2025, Z-2656-2025, Z-2657-2025, Z-2658-2025, Z-2659-2025, Z-2660-2025, Z-2661-2025, Z-2662-2025, Z-2663-2025, Z-2664-2025, Z-2665-2025 and 3 more
Recalling firm Siemens Medical Solutions USA, Inc
Product type Devices
Products Biograph mMR. Model Number: 10433372., BIOGRAPH One (DE). Model Number: 11689172., MAGNETOM Cima.X (DE). Model Number: 11647158., MAGNETOM Connectom.X. Model Number: 11371480., MAGNETOM Lumina (DE). Model Number: 11344916. and 10 more
Classification Class I
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths…
Initiation date Aug 28, 2025
Report date Oct 8, 2025
City Malvern
State PA
Country United States
FEI number 2240869

At a glance

  • 10–99 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls