Drug recall

UniCel DxH 600 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF:B23858

FDA recall event 97464 · Class II · Beckman Coulter Inc.

Key facts

Event ID 97464
Recall numbers Z-2572-2025, Z-2573-2025, Z-2574-2025, Z-2575-2025
Recalling firm Beckman Coulter Inc.
Product type Devices
Products UniCel DxH 600 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF:B23858, UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520, UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322, UniCel DxH 900 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C11478
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
Initiation date Aug 13, 2025
Report date Sep 24, 2025
City Brea
State CA
Country United States
FEI number 2050012

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: