Drug recall

MEDLINE ReNewal Coblator II ENT Evac 70 Xtra HP,w/Integrated Cable (Blue/Grey), Number REF EICA58740R

FDA recall event 96541 · Class II · MEDLINE INDUSTRIES, LP - Northfield

Key facts

Event ID 96541
Recall numbers Z-1546-2025, Z-1547-2025, Z-1548-2025, Z-1549-2025
Recalling firm MEDLINE INDUSTRIES, LP - Northfield
Product type Devices
Products MEDLINE ReNewal Coblator II ENT Evac 70 Xtra HP,w/Integrated Cable (Blue/Grey), Number REF EICA58740R, MEDLINE ReNewal Coblator II Evac 70 Xtra Plasma Wand,w/Integrated Saline Line & Integrated Cable, Item Number REF EICA58720R, MEDLINE ReNewal ENT Coblator II PROcise mAx,W/ Integrated Cable Suction & Saline (Blue), Item Number REF EICA88980R, MEDLINE ReNewal ENT Coblator II PROcise XP Wand,w/Integrated Cable Suction & Saline (Blue), Item Number REF EICA88720R
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Medline ReNewal has identified that the drip chambers of affected devices may disconnect from the saline line tubing when under tension.
Initiation date Mar 5, 2025
Report date Apr 16, 2025
City Northfield
State IL
Country United States
FEI number 1417592

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls