Drug recall
IntelliBridge EC10 Module. Model Number: 865115
FDA recall event 96138 · Class II · Philips North America Llc
Key facts
| Event ID | 96138 |
|---|---|
| Recall numbers | Z-1106-2025, Z-1107-2025, Z-1108-2025, Z-1109-2025, Z-1110-2025 |
| Recalling firm | Philips North America Llc |
| Product type | Devices |
| Products | IntelliBridge EC10 Module. Model Number: 865115, IntelliVue Patient Monitor MX400. Model Number: 866060, IntelliVue Patient Monitor MX450. Model Number: 866062, IntelliVue Patient Monitor MX500. Model Number: 866064, IntelliVue Patient Monitor MX550. Model Number: 866066 |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | Potential failure to alarm for "No Device Data" INOP. The device to alarm after a Hamilton ventilator is disconnected from the EC10 Module or IntelliBridge I/O board. |
| Initiation date | Jan 15, 2025 |
| Report date | Feb 12, 2025 |
| City | Cambridge |
| State | MA |
| Country | United States |
| FEI number | 3016618143 |