Drug recall
Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product…
FDA recall event 96107 · Class II · Sysmex America, Inc.
Key facts
| Event ID | 96107 |
|---|---|
| Recall numbers | Z-1113-2025 |
| Recalling firm | Sysmex America, Inc. |
| Product type | Devices |
| Products | Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow… |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System. |
| Initiation date | Jan 6, 2025 |
| Report date | Feb 12, 2025 |
| City | Lincolnshire |
| State | IL |
| Country | United States |
| FEI number | 3009711478 |