Drug recall

Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product…

FDA recall event 96107 · Class II · Sysmex America, Inc.

Key facts

Event ID 96107
Recall numbers Z-1113-2025
Recalling firm Sysmex America, Inc.
Product type Devices
Products Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow…
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.
Initiation date Jan 6, 2025
Report date Feb 12, 2025
City Lincolnshire
State IL
Country United States
FEI number 3009711478

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: