Drug recall

cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in…

FDA recall event 95609 · Class II · Roche Diagnostics Operations, Inc.

Key facts

Event ID 95609
Recall numbers Z-0489-2025
Recalling firm Roche Diagnostics Operations, Inc.
Product type Devices
Products cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total L-homocysteine
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Abnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)
Initiation date Oct 17, 2024
Report date Nov 27, 2024
City Indianapolis
State IN
Country United States
FEI number 1823260

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls