Drug recall

AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood…

FDA recall event 95497 · Class II · Accriva Diagnostics, Inc.

Key facts

Event ID 95497
Recall numbers Z-0279-2025
Recalling firm Accriva Diagnostics, Inc.
Product type Devices
Products AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Values used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results…
Initiation date Sep 27, 2024
Report date Nov 6, 2024
City San Diego
State CA
Country United States
FEI number 2250033

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls