Drug recall
Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing…
FDA recall event 95472 · Class II · Zimmer, Inc.
Key facts
| Event ID | 95472 |
|---|---|
| Recall numbers | Z-0285-2025 |
| Recalling firm | Zimmer, Inc. |
| Product type | Devices |
| Products | Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing… |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into… |
| Initiation date | Sep 17, 2024 |
| Report date | Nov 13, 2024 |
| City | Warsaw |
| State | IN |
| Country | United States |
| FEI number | 1000220733 |