Drug recall

Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing…

FDA recall event 95472 · Class II · Zimmer, Inc.

Key facts

Event ID 95472
Recall numbers Z-0285-2025
Recalling firm Zimmer, Inc.
Product type Devices
Products Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing…
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into…
Initiation date Sep 17, 2024
Report date Nov 13, 2024
City Warsaw
State IN
Country United States
FEI number 1000220733

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls