Drug recall

AET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock…

FDA recall event 95153 · Class II · B. Braun Medical, Inc.

Key facts

Event ID 95153
Recall numbers Z-3144-2024
Recalling firm B. Braun Medical, Inc.
Product type Devices
Products AET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock, Model/Catalog Number: 472036
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Potential for product leakage from the Spin Lock Connector on the distal end of the device due to the migration of excess solvent during the assembly process.
Initiation date Aug 8, 2024
Report date Sep 25, 2024
City Allentown
State PA
Country United States
FEI number 2523676

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls