Drug recall

Philips Azurion 7M20 systems with FlexArm ceiling-mounted system, Interventional Fluoroscopic X-Ray System…

FDA recall event 95148 · Class II · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Key facts

Event ID 95148
Recall numbers Z-2953-2024
Recalling firm PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product type Devices
Products Philips Azurion 7M20 systems with FlexArm ceiling-mounted system, Interventional Fluoroscopic X-Ray System, Model Numbers 722079 and 722224
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or…
Initiation date Aug 6, 2024
Report date Sep 11, 2024
City Best
Country Netherlands
FEI number 3002807880

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls