Drug recall

Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D…

FDA recall event 95017 · Class II · Palodex Group Oy

Key facts

Event ID 95017
Recall numbers Z-2413-2024
Recalling firm Palodex Group Oy
Product type Devices
Products Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device is…
Classification Class II
Status Ongoing
Voluntary or mandated FDA Requested
Reason for recall It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria…
Initiation date Oct 28, 2021
Report date Aug 21, 2024
City Tuusula
Country Finland
FEI number 3005383085

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls