Drug recall
Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT, NA/6000-00-00; Arctic Sun STAT Loaner…
FDA recall event 94881 · Class II · Medivance Inc.
Key facts
| Event ID | 94881 |
|---|---|
| Recall numbers | Z-2457-2024 |
| Recalling firm | Medivance Inc. |
| Product type | Devices |
| Products | Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT, NA/6000-00-00; Arctic Sun STAT Loaner, NA/6000-00-00L; Arctic Sun Stat EU UK/600001; Arctic Sun Stat Taiwan/600006; Arctic Sun… |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | If temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden patient temperature changes/interruption in water… |
| Initiation date | Jun 27, 2024 |
| Report date | Aug 7, 2024 |
| City | Louisville |
| State | CO |
| Country | United States |
| FEI number | 3003329224 |