Drug recall

MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504…

FDA recall event 94609 · Class II · Canon Medical System, USA, INC.

Key facts

Event ID 94609
Recall numbers Z-2041-2024
Recalling firm Canon Medical System, USA, INC.
Product type Devices
Products MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage Titan 3T MRT-3010 (MEXL-3010), Vantage Galan 3T MRT-3020 (MEXL-3020)
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing…
Initiation date May 10, 2024
Report date Jun 19, 2024
City Tustin
State CA
Country United States
FEI number 2020563