Drug recall

DxI 9000 Access Immunoassay Analyzer, C11137; in-vitro diagnostic

FDA recall event 94499 · Class II · Beckman Coulter, Inc.

Key facts

Event ID 94499
Recall numbers Z-2264-2024
Recalling firm Beckman Coulter, Inc.
Product type Devices
Products DxI 9000 Access Immunoassay Analyzer, C11137; in-vitro diagnostic
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Beckman Coulter is recalling their DxI 9000 Access Immunoassay Analyzers, an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte…
Initiation date Apr 18, 2024
Report date Jul 10, 2024
City Chaska
State MN
Country United States
FEI number 2122870

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls