Drug recall
Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum…
FDA recall event 94358 · Class II · Randox Laboratories Ltd.
Key facts
| Event ID | 94358 |
|---|---|
| Recall numbers | Z-1759-2024 |
| Recalling firm | Randox Laboratories Ltd. |
| Product type | Devices |
| Products | Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinical chemistry tests. |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods. |
| Initiation date | Mar 5, 2024 |
| Report date | May 15, 2024 |
| City | Crumlin (North) |
| Country | Ireland |
| FEI number | 2965 |