Drug recall

Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee…

FDA recall event 94299 · Class II · Siemens Medical Solutions USA, Inc

Key facts

Event ID 94299
Recall numbers Z-1598-2024
Recalling firm Siemens Medical Solutions USA, Inc
Product type Devices
Products Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego…
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE…
Initiation date Mar 21, 2024
Report date Apr 24, 2024
City Malvern
State PA
Country United States
FEI number 2240869

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls