Drug recall

Synapse PACS - Version 7.1.000

FDA recall event 94090 · Class II · FUJIFILM Healthcare Americas Corporation

Key facts

Event ID 94090
Recall numbers Z-1268-2024, Z-1269-2024, Z-1270-2024, Z-1271-2024, Z-1272-2024, Z-1273-2024
Recalling firm FUJIFILM Healthcare Americas Corporation
Product type Devices
Products Synapse PACS - Version 7.1.000, Synapse PACS - Version 7.1.000US, Synapse PACS - Version 7.2.000, Synapse PACS - Version 7.2.100, Synapse PACS - Version 7.2.200 and 1 more
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
Initiation date Jan 10, 2024
Report date Mar 13, 2024
City Lexington
State MA
Country United States
FEI number 1000513161

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls