Drug recall

Access Total T4, REF 33800, IVD, thyroxine reagent assay.

FDA recall event 94012 · Class II · Beckman Coulter, Inc.

Key facts

Event ID 94012
Recall numbers Z-1303-2024
Recalling firm Beckman Coulter, Inc.
Product type Devices
Products Access Total T4, REF 33800, IVD, thyroxine reagent assay.
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.
Initiation date Feb 8, 2024
Report date Mar 27, 2024
City Chaska
State MN
Country United States
FEI number 2122870

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: