Drug recall

Achieva 1.5T Conversion. Model (REF) Numbers 781346, 781283.

FDA recall event 93787 · Class II · Philips North America

Key facts

Event ID 93787
Recall numbers Z-0945-2024, Z-0946-2024, Z-0947-2024, Z-0948-2024, Z-0949-2024, Z-0950-2024, Z-0951-2024, Z-0952-2024, Z-0953-2024, Z-0954-2024, Z-0955-2024, Z-0956-2024
Recalling firm Philips North America
Product type Devices
Products Achieva 1.5T Conversion. Model (REF) Numbers 781346, 781283., Achieva 1.5T Initial system. Model (REF) Numbers 781178., Achieva 1.5T. Model (REF) Numbers 781196, 781343, 781296., Achieva 3.0T. Model. (REF) Numbers 781277, 781177, 781278, 781344, 781345., Achieva XR. Model (REF) Numbers 781153, 781253. and 7 more
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient…
Initiation date Jan 3, 2024
Report date Feb 7, 2024
City Cambridge
State MA
Country United States
FEI number 3006648320

At a glance

  • 10–99 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: