Drug recall

Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)

FDA recall event 93734 · Class II · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Key facts

Event ID 93734
Recall numbers Z-0861-2024
Recalling firm PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product type Devices
Products Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.
Initiation date Dec 8, 2023
Report date Feb 7, 2024
City Best
Country Netherlands
FEI number 3002807880

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls