Drug recall

Medtronic SynchroMed II, Model 8637-20, Programmable pump

FDA recall event 93481 · Class II · Medtronic Neuromodulation

Key facts

Event ID 93481
Recall numbers Z-0614-2024, Z-0615-2024
Recalling firm Medtronic Neuromodulation
Product type Devices
Products Medtronic SynchroMed II, Model 8637-20, Programmable pump, Medtronic SynchroMed II, Model 8637-40, Programmable pump
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will…
Initiation date Nov 13, 2023
Report date Jan 3, 2024
City Minneapolis
State MN
Country United States
FEI number 2182207

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls