Drug recall

Novum IQ Syringe infusion system, Product Code 40800BAXUS

FDA recall event 93191 · Class I · Baxter Healthcare Corporation

Key facts

Event ID 93191
Recall numbers Z-0151-2024
Recalling firm Baxter Healthcare Corporation
Product type Devices
Products Novum IQ Syringe infusion system, Product Code 40800BAXUS
Classification Class I
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Baxter is issuing an Urgent Medical Device Correction for the Novum IQ Syringe Pump. Baxter identified that after multiple downstream occlusion alarms, the pump may display an Infusion Complete alarm…
Initiation date Oct 13, 2023
Report date Nov 15, 2023
City Deerfield
State IL
Country United States
FEI number 1417572

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls