Drug recall

Puritan Bennett Cuff Pressure Manager, REF: 180-03

FDA recall event 92941 · Class II · Covidien

Key facts

Event ID 92941
Recall numbers Z-2651-2023
Recalling firm Covidien
Product type Devices
Products Puritan Bennett Cuff Pressure Manager, REF: 180-03
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which…
Initiation date Aug 22, 2023
Report date Oct 11, 2023
City Boulder
State CO
Country United States
FEI number 2936999

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: