Drug recall

Link OptiStem Rasp Handle-Used with the OptiStem Rasp to prepare the femoral medullary canal for a knee…

FDA recall event 92839 · Class II · Waldemar Link GmbH & Co. KG (Mfg Site)

Key facts

Event ID 92839
Recall numbers Z-2481-2023
Recalling firm Waldemar Link GmbH & Co. KG (Mfg Site)
Product type Devices
Products Link OptiStem Rasp Handle-Used with the OptiStem Rasp to prepare the femoral medullary canal for a knee prosthesis implant during a total knee arthroplasty surgical procedure Item Number: 17-5220/01
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall A crack in the weld of the rasp handle may lead to a fracture of the frontal sleeve when the rasp is further impacted or extracted. The rasp remains intact. The rasp would need to be pulled out by a…
Initiation date Jul 26, 2023
Report date Sep 6, 2023
City Norderstedt
Country Germany
FEI number 3003386935

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: