Drug recall

Ingenia Elition S Magnetic Resonance (MR) system, Product Numbers 781357 and 782106

FDA recall event 92720 · Class II · Philips North America

Key facts

Event ID 92720
Recall numbers Z-2395-2023, Z-2396-2023, Z-2397-2023
Recalling firm Philips North America
Product type Devices
Products Ingenia Elition S Magnetic Resonance (MR) system, Product Numbers 781357 and 782106, Ingenia Elition X Magnetic Resonance (MR) system, Product Numbers 781358, 782107, and 782118 (SmartPath to Ingenia Elition X), MR 7700 Magnetic Resonance (MR) system, Product Numbers 782120 and 782130 (Upgrade to MR 7700)
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.
Initiation date Jun 12, 2023
Report date Aug 16, 2023
City Cambridge
State MA
Country United States
FEI number 3006648320

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls